User Manual 1178.8520.02 ─ 01
Risk of rendering instrument unusable
The instrument is equipped with the Windows
10 operating system. You can install
additional software on the instrument, however, additional software can impair instru-
ment function. Thus, run only programs that Rohde
&
Schwarz has tested for compati-
bility with the instrument software.
The drivers and programs used on the instrument under Windows
10 are adapted to
the instrument. Only install update software released by Rohde
&
Schwarz to modify
existing instrument software.
The following program packages have been tested:
●
R&S Power Viewer Plus - virtual power meter for displaying results of the power
sensor R&S NRP-Zxx (install only this component!)
●
Symantec Endpoint Security – virus-protection software
●
FileShredder - for reliable deletion of files on the hard disk
5.1.2.1
Virus Protection
Take appropriate steps to protect your instruments from infection. Use strong firewall
settings and scan any removable storage device used with a Rohde
&
Schwarz instru-
ment regularly. It is also recommended that you install anti-virus software on the instru-
ment. Rohde
&
Schwarz does NOT recommend running anti-virus software in the
background ("on-access" mode) on Windows-based instruments, due to potentially
degrading instrument performance. However, Rohde
&
Schwarz does recommend run-
ning it during non-critical hours.
For details and recommendations, see the following Rohde
&
Schwarz white paper:
●
1EF96: Malware Protection Windows 10
5.1.2.2
Service Packs and Updates
Microsoft regularly creates security updates and other patches to protect Windows-
based operating systems. These are released through the Microsoft Update website
and associated update server. Instruments using Windows, especially those that con-
nect to a network, should be updated regularly.
For details and recommendations, see the following Rohde
&
Schwarz white paper:
●
1EF96: Malware Protection Windows 10
5.1.2.3
Login
Windows
10 requires that users identify themselves by entering a user name and pass-
word in a login window. By default, the R&S
FSV/A provides two user accounts:
●
"Instrument"
: a standard user account with limited access