Common Measurement Settings
R&S
®
FSVA3000/ R&S
®
FSV3000
326
User Manual 1178.8520.02 ─ 01
External Generator Connections
The external generator is controlled either via a LAN connection or via the EXT. GEN.
CONTROL GPIB interface of the R&S
FSV/A supplied with the option.
For more information on configuring interfaces see
Chapter 13.1.1, "Remote Control
TTL synchronization
In addition, TTL synchronization can be used with some Rohde
&
Schwarz generators
connected via GPIB. The TTL interface is included in the AUX control connector of the
External Generator Control option.
Using the TTL interface allows for considerably higher measurement rates than pure
GPIB control, because the frequency stepping of the R&S
FSV/A is directly coupled
with the frequency stepping of the generator. For details see
In
, the connection for an R&S SMW is shown.
R&S SMW
rear panel
R&S FSW
rear panel
BNC Blank
BNC Trigger
Figure 8-3: TTL connection for an R&S SMW generator
Data Input and Output