Common Measurement Settings
R&S
®
FSVA3000/ R&S
®
FSV3000
336
User Manual 1178.8520.02 ─ 01
Overloading
At a reference level of -10 dBm and at an external generator output level of the same
value, the R&S
FSV/A operates without overrange reserve. That means the
R&S
FSV/A is in danger of being overloaded if a signal is applied whose amplitude is
higher than the reference line. In this case, either the message "RF OVLD" for over-
load or "ADC OVLD" for exceeded display range (clipping of the trace at the upper dia-
gram border = overrange) is displayed in the status line.
Overloading can be avoided as follows:
●
Reducing the output level of the external generator (
in "External Generator > Measurement Configuration")
●
Increasing the reference level (
in the "Amplitude" menu)
8.2.4.3
External Generator Control Settings
Access
: [INPUT/OUPUT] > "External Generator Config"
The "External Generator" settings are available if the R&S
FSV/A External Generator
Control option is installed. For each measurement channel, you can configure one
external generator. To switch between different configurations, define multiple mea-
surement channels.
For more information on external generator control, see
Interface Configuration Settings
............................................................................336
.......................................................................................... 338
................................................................................341
Interface Configuration Settings
Access
: [INPUT/OUPUT] > "External Generator Config" > "Interface Configuration" tab
Data Input and Output