Defines the usage of the variable Trigger Input/Output connectors, where:
"Trigger 2" : Trigger Input/Output connector on the front panel
"Trigger 3" : Trigger 3 Input/Output connector on the rear panel
Defines the usage of the variable Trigger Aux connector on the rear panel.
(Trigger 1 is INPUT only.)
"Input"
The signal at the connector is used as an external trigger source by
the R&S
FSV/A. Trigger input parameters are available in the "Trig-
ger" dialog box.
"Output"
The R&S
FSV/A sends a trigger signal to the output connector to be
used by connected devices.
Further trigger parameters are available for the connector.
Remote command:
OUTPut<up>:TRIGger<tp>:DIRection
Output Type ← Trigger 1/2
Type of signal to be sent to the output
"Trigger Off"
Deactivates the output. (Only for Trigger 3, for which only output is
supported.)
"Device Trig-
gered"
(Default) Sends a trigger when the R&S
FSV/A triggers.
"Trigger
Armed"
Sends a (high level) trigger when the R&S
FSV/A is in "Ready for trig-
ger" state.
This state is indicated by a status bit in the
STATus:OPERation
reg-
ister (bit 5), as well as by a low-level signal at the AUX port (pin 9).
For details, see
R&S
FSV/A Getting Started manual.
Data Input and Output