Common Measurement Settings
R&S
®
FSVA3000/ R&S
®
FSV3000
386
User Manual 1178.8520.02 ─ 01
Remote command:
[SENSe:]BANDwidth[:RESolution]:TYPE
Default Coupling
Sets all coupled functions to the default state ( "Auto" ). In addition, the ratio "RBW/
VBW" is set to "Sine[1/1]" and the ratio "Span/RBW" to
100.
For more information see
Chapter 8.5.1.3, "Coupling VBW and RBW"
Remote command:
[SENSe:]BANDwidth[:RESolution]:AUTO
Sweep/Average Count
Defines the number of sweeps to be performed in the single sweep mode. Values from
0 to 200000 are allowed. If the values 0 or 1 are set, one sweep is performed.
The sweep count is applied to all the traces in all diagrams.
If the trace modes "Average" , "Max Hold" or "Min Hold" are set, this value also deter-
mines the number of averaging or maximum search procedures.
In continuous sweep mode, if "Sweep Count" = 0 (default), averaging is performed
over 10 sweeps. For "Sweep Count" =1, no averaging, maxhold or minhold operations
are performed.
For more information, see
Chapter 8.5.1.8, "How Much Data is Measured: Sweep
Remote command:
Sweep Points
Defines the number of measured values to be collected during one sweep.
For details see
Chapter 8.5.1.8, "How Much Data is Measured: Sweep Points and
Remote command:
[SENSe:]SWEep[:WINDow<n>]:POINts
Optimization
In FFT mode, several FFT analysis steps are required to cover the entire measurement
span. The span which is covered by one FFT analysis step is called
partial span
. The
partial span cannot be defined directly, but it can be optimized according to measure-
ment requirements.
Bandwidth, Filter and Sweep Configuration