Common Measurement Settings
R&S
®
FSVA3000/ R&S
®
FSV3000
388
User Manual 1178.8520.02 ─ 01
Remote command:
CALCulate<n>:SPECtrogram:CONTinuous
Continuous Sweep / Run Cont
After triggering, starts the sweep and repeats it continuously until stopped. This is the
default setting.
After triggering, starts the measurement and repeats it continuously until stopped.
While the measurement is running, the "Continuous Sweep" softkey and the [RUN
CONT] key are highlighted. The running measurement can be aborted by selecting the
highlighted softkey or key again. The results are not deleted until a new measurement
is started.
Note:
Sequencer. If the Sequencer is active, the "Continuous Sweep" softkey only con-
trols the sweep mode for the currently selected channel. However, the sweep mode
only takes effect the next time the Sequencer activates that channel, and only for a
channel-defined sequence. In this case, a channel in continuous sweep mode is swept
repeatedly.
Furthermore, the [RUN CONT] key controls the Sequencer, not individual sweeps.
[RUN CONT] starts the Sequencer in continuous mode.
For details on the Sequencer, see
Chapter 6.4.1, "The Sequencer Concept"
Remote command:
Continue Single Sweep
After triggering, repeats the number of sweeps set in "Sweep Count", without deleting
the trace of the last measurement.
While the measurement is running, the "Continue Single Sweep" softkey and the [RUN
SINGLE] key are highlighted. The running measurement can be aborted by selecting
the highlighted softkey or key again.
Remote command:
Spectrogram Frames
These settings are only available if spectrogram display is active.
For more information see
Chapter 9.5.2.3, "How to Display and Configure a Spectro-
Select Frame ← Spectrogram Frames
Selects a specific frame, loads the corresponding trace from the memory, and displays
it in the Spectrum window.
Note that activating a marker or changing the position of the active marker automati-
cally selects the frame that belongs to that marker.
This function is only available in single sweep mode or if the sweep is stopped, and
only if a spectrogram is selected.
The most recent frame is number 0, all previous frames have a negative number.
Bandwidth, Filter and Sweep Configuration