Common Analysis and Display Functions
R&S
®
FSVA3000/ R&S
®
FSV3000
447
User Manual 1178.8520.02 ─ 01
CALCulate<n>:MARKer<m>:FUNCtion:NOISe:RESult?
..........................................................................................447
Switching All Noise Measurement Off
........................................................................447
Marker State
Activates or deactivates the marker in the diagram.
Remote command:
CALCulate<n>:MARKer<m>[:STATe]
CALCulate<n>:DELTamarker<m>[:STATe]
Marker Type
Toggles the marker type.
The type for marker 1 is always "Normal" , the type for delta marker 1 is always
"Delta" . These types cannot be changed.
Note:
If normal marker 1 is the active marker, switching the "Mkr Type" activates an
additional delta marker 1. For any other marker, switching the marker type does not
activate an additional marker, it only switches the type of the selected marker.
"Normal"
A normal marker indicates the absolute value at the defined position
in the diagram.
"Delta"
A delta marker defines the value of the marker relative to the speci-
fied reference marker (marker 1 by default).
Remote command:
CALCulate<n>:MARKer<m>[:STATe]
CALCulate<n>:DELTamarker<m>[:STATe]
Noise Measurement State
Activates or deactivates noise measurement for the marker in the diagram.
This function is only available for normal markers.
If activated, the marker displays the noise power density measured at the position of
the marker.
For details see
Chapter 9.3.4.2, "Measuring Noise Density (Noise Meas Marker)"
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:NOISe[:STATe]
CALCulate<n>:MARKer<m>:FUNCtion:NOISe:RESult?
Switching All Noise Measurement Off
Deactivates noise measurement for all markers.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:NOISe[:STATe]
Marker Usage