Common Analysis and Display Functions
R&S
®
FSVA3000/ R&S
®
FSV3000
457
User Manual 1178.8520.02 ─ 01
audio output. Thus, a displayed signal can be identified acoustically with the help of the
internal loudspeaker or with headphones.
This function is not available for Spectrum Emission Mask measurements or measure-
ments on I/Q-based data.
The sweep stops at the frequency determined by marker 1 for the selected time and
the RF signal is demodulated in a bandwidth that corresponds to the RBW. Alterna-
tively, demodulation can be activated continuously, i.e. audio output occurs regardless
of the marker position and the marker stop time. For measurements in the time domain
(zero span), demodulation is always continuous.
Optionally, a minimum level ( "Squelch Level" ) can be defined so that the signal is only
demodulated when it exceeds the set level. This is useful during continuous demodula-
tion to avoid listening to noise.
The squelch function activates the video trigger function (see
and deactivates any other trigger or gating settings. The squelch level and trigger level
are set to the same value. The trigger source in the channel bar is indicated as "SQL"
for squelch. The squelch level is indicated by a red line in the diagram.
Remote commands:
"Examples: Demodulating Marker Values and Providing Audio Output"
....................................................................................... 458
..........................................................................................458
Marker Usage