Data Management
R&S
®
FSVA3000/ R&S
®
FSV3000
532
User Manual 1178.8520.02 ─ 01
Saving instrument settings in secure user mode
Be sure to store instrument settings that you require beyond the current session before
Chapter 5.1.7, "Protecting Data Using the Secure
Settings that are saved via QuickSave in secure user mode are only available during
the current session. As soon as the power is switched off on the R&S
FSV/A, the data
is cleared.
Saving and recalling transducer and limit line settings
If a transducer file was in use when the save set was stored (with the save item "Cur-
rent Settings" only) the R&S
FSV/A assumes that these transducer values should
remain valid after every recall of that save set. Thus, even if the transducer file is
changed and the original save set file is recalled later, the
originally stored
transducer
values are recalled and applied to the measurement. In the "Edit" transducer dialog
box, however, the
changed
transducer file values are displayed, as no updated trans-
ducer file was loaded.
The same applies to limit line settings.
The same applies to integrated measurements' weighting filter.
If you want to apply the changed transducer values after recalling the save set, you
must force the application to reload the transducer file. To do so, simply open the "Edit
Transducer" dialog box (see
Chapter 12.5.2, "Transducer Settings"
and toggle the "X-Axis" option from "Lin" to "log" and back. Due to that change, the
transducer file is automatically reloaded, and the changed transducer values are
applied to the current measurement. Now you can create a new save set with the
updated transducer values.
Similarly, if you want to apply the changed limit values after recalling the save set, you
must force the application to reload the limit file. To do so, simply open the "Edit Limit
Line" dialog box (see
Chapter 9.4.2.2, "Limit Line Settings and Functions"
on page 474) and toggle the "Y-Axis" unit. Due to that change, the limit line file is auto-
matically reloaded, and the changed limit values are applied to the current measure-
ment. Now a new save set with the updated limit values can be created.
.......................................................................................532
Configurable Storage and Recall
.......................................................................... 534
How to Save and Load Instrument Settings
..........................................................540
11.3.1
Quick Save/Quick Recall
The Quick Save and Quick Recall functions allow you to store instrument settings or
channels very easily and quickly in one step. Up to ten different sets of settings can be
stored to or recalled from "save sets". Each save set is identified by its storage date
and type (instrument or specific "Channel" ) in the display. The save sets are stored in
the
C:\R_S\INSTR\USER\QuickSave
directory, in files named
QuickSave1.dfl
to
QuickSave10.dfl
. The storage filenames and locations cannot be changed. Only
the current measurement settings are stored, not any additional data such as traces,
Storing and Recalling Instrument Settings and Measurement Data