Getting Started
R&S
®
FSVA3000/ R&S
®
FSV3000
55
User Manual 1178.8520.02 ─ 01
All USB connectors support standard 2.0.
5.2.2.5
Aux. Port
The 9-pole SUB-D male connector provides control signals for controlling external
devices. The voltage levels are of the TTL type (max. 5
V).
Short-circuit hazard
Always observe the designated pin assignment. A short-circuit can damage the port.
5.2.2.6
Aux. Control
The 9-pole SUB-D female connector allows the R&S
FSV/A to receive control signals
from a controlling external device, such as an external generator. The voltage levels
are of the TTL type (max. 5
V).
Short-circuit hazard
Always observe the designated pin assignment. A short-circuit can damage the port.
5.2.2.7
Trigger 2/3 In/Out
The additional female BNC Trigger 2 In/Out connector and the optional Trigger 3 (out-
put) connector allow the R&S
FSV/A to receive additional external signals or to provide
signals to another device. The signals are TTL compatible (0
V / 5
V). You control the
connector usage in the "Trigger" settings ([Trigger] key).
5.2.2.8
Video Output
The female BNC connector can be used to provide video output (1 V).
5.2.2.9
IF Output
The female BNC connector can be used to output the intermediate frequency (IF).