General Instrument Setup
R&S
®
FSVA3000/ R&S
®
FSV3000
564
User Manual 1178.8520.02 ─ 01
Self-alignment results in secure user mode
Be sure to store self-alignment results before
ter 5.1.7, "Protecting Data Using the Secure User Mode"
In secure user mode, the results are not stored permanently. Thus, if the currently
stored self-alignment results are not suitable, you must perform a self-alignment each
time you switch on the R&S
FSV/A.
Await Warm-Up Operation before Self Alignment
....................................................... 565
Shut down Device after Self Alignment
.......................................................................565
..........................................................................................565
Starting Touch Screen Alignment
...............................................................................565
Start Self Alignment
Starts recording correction data for the instrument. If the correction data acquisition
fails or if the correction values are deactivated, a corresponding message is displayed
in the status field.
For details, see
Chapter 12.1.1, "Basics on Alignment"
Note:
A running Sequencer operation is aborted when you start a self-alignment.
Remote command:
Abort Self Alignment
As long as the self-alignment data is being collected, the procedure can be canceled
using the "Abort Self Alignment" button.
Alignment