General Instrument Setup
R&S
®
FSVA3000/ R&S
®
FSV3000
600
User Manual 1178.8520.02 ─ 01
Table 12-1: ASCII file format for transducer factor files
File contents
Description
Header data
sep=;
Separator for individual values (required by Microsoft Excel, for
example)
Type
;RS_TransducerFactor;
Type of data
FileFormatVersion
;1.00;
File format version
Date;01.Oct 2006;
Date of data set storage
OptionID;SpectrumAnalyzer
Application the transducer factor was created for
Name;
TestTDF1
Transducer factor name
Comment;Transducer for device A
Description of transducer factor
XAxisScaling;LINEAR
Scaling of x-axis linear (LIN) or logarithmic (LOG)
YAxisUnit;LEVEL_DB
Unit of y values
YAxisScaleMode;ABSOLUTE
Scaling of y-axis (absolute or relative)
NoOfPoints
;5
Number of points the line is defined by
Data section for individual data points
100000000;-50.000000
x- and y-values of each data point defining the line
500000000;-30.000000
1000000000;0.000000
1500000000;-30.000000
2500000000;-50.000000
12.5.4
How to Configure the Transducer
Configuring the transducer is very similar to configuring transducer factors.
The transducer settings are defined in the "Transducer" dialog box which is displayed
when you press the [Setup] key and then select "Transducer" .
Stored transducer settings
When storing and recalling transducer settings, consider the information provided in
"Saving and recalling transducer and limit line settings"
The following tasks are described:
●
"How to find compatible transducer lines"
●
"How to activate and deactivate a transducer"
●
"How to edit existing transducer lines"
●
"How to copy an existing transducer line"
●
"How to delete an existing transducer line"
Transducers