If a measurement is running, stop it by pressing the highlighted [Run Cont] or [Run
Single] key.
3. Select the [Setup] key.
4. Select the "System Config" softkey.
5. Select the "Firmware Update" tab.
6. In the file selection dialog box select the
FSV3000Setup*.exe
file.
7. Select "Install" to start the update.
8. After the firmware update, the R&S
FSV/A reboots automatically.
9. Depending on the previous firmware version, a reconfiguration of the hardware
might be required during the first startup of the firmware. The reconfiguration starts
automatically, and a message box informs you about the process. When the recon-
figuration has finished, the instrument again reboots automatically.
Note:
Do not switch off the instrument during the reconfiguration process!
Now the firmware update is complete.
It is recommended that you perform a self-alignment after the update (see
ter 12.1.4, "How to Align the Instrument"
12.8.5
General Configuration Settings
Access
: [Setup] > "System Configuration" > "Config"
General system settings, for example concerning the initial behaviour of the
R&S
FSV/A after booting, can also be configured.
...................................................................................... 622
................................................................................ 623
............................................................................................... 624
Out-of-range value behavior
By default, if you enter a value that is outside the valid range in an input field for a set-
ting, a warning is displayed and the value is not accepted. Alternatively, entries below
the minimum value can automatically be set to the minimum possible entry, and entries
above the maximum value set to the maximum possible entry. This behavior avoids
errors and facilitates setting correct values.
System Configuration Settings