General Instrument Setup
R&S
®
FSVA3000/ R&S
®
FSV3000
642
User Manual 1178.8520.02 ─ 01
Synchronizing all Spectrum channels
The following example demonstrates how to synchronize the center frequency in all
Spectrum application channels, while the VSA and Analog Demodulation applications
remain independent.
1. Select the [SETUP] key.
2. Select "Parameter Coupling".
3. Select the "User Coupling" tab.
4. Select "Add New User Coupling".
5. From the "Channel 1" list, select "All Spectrum".
6. From the "Coupling Element 1" list, select "Center Frequency".
7. Select "Couple Selected Parameters".
8. Close the "Parameter Coupling" dialog box.
9. In the first Spectrum channel, change the "Center Frequency" to 1
GHz.
10. Switch to the second Spectrum channel.
The center frequency in the second Spectrum channel is also set to 1
GHz.
Synchronizing specific channels
The following example demonstrates how to synchronize the attenuation only for the
first Spectrum channel and the first Analog Demodulation channel, while the other
three channels remain independent.
1. Select "Add New User Coupling".
2. From the "Channel 1" list, select "Spectrum 1".
3. From the "Coupling Element 1" list, select "Attenuation".
4. From the "Channel 2" list, select "AnaDemod 1".
5. From the "Coupling Element 2" list, select "Attenuation".
6. Select "Couple Selected Parameters".
7. Close the "Parameter Coupling" dialog box.
8. In the first Spectrum channel, change the "Attenuation" to 15
dB.
9. Switch to the first Analog Demodulation channel.
The attenuation in the second Analog Demodulation channel is also set to 15
dB.
Synchronizing Measurement Channel Configuration