set the hardware explicitly. In any case, the hardware is set automatically when a
sweep is performed.
Reacting to malfunctions
The service request is the only possibility for the instrument to become active on its
own. Each controller program should instruct the instrument to initiate a service
request in case of malfunction. The program should react appropriately to the service
request.
Error queues
The error queue should be queried after every service request in the controller pro-
gram as the entries describe the cause of an error more precisely than the status regis-
ters. Especially in the test phase of a controller program the error queue should be
queried regularly since faulty commands from the controller to the instrument are
recorded there as well.
13.2
GPIB Languages
The R&S
FSV/A analyzer family supports a subset of the GPIB commands used by
other devices. Thus it can emulate other devices in order to use existing remote control
programs.
The device model to be emulated is selected manually using "SETUP > N
Remote > GPIB tab > Language". Via the GPIB interface using the
In order to emulate device models that are not part of the selection list of the GPIB
"Language" setting, you can modify the identification string received in response to the
ID command ("Identification String" setting). Thus, any device model whose command
set is compatible with one of the supported device models can be emulated.
Supported languages
Language
Comment
SCPI
71100C
Compatible to 8566A/B
71200C
Compatible to 8566A/B
71209A
Compatible to 8566A/B
8560E
8561E
8562E
8563E
8564E
GPIB Languages