Network and Remote Operation
R&S
®
FSVA3000/ R&S
®
FSV3000
701
User Manual 1178.8520.02 ─ 01
Remote Display Update
Defines whether the display of the R&S
FSV/A is updated when changing from manual
operation to remote control.
Turning off the display update function improves performance during remote control.
Note:
Usually, this function remains available on the display during remote operation.
However, it can be disabled remotely. In this case, the display is not updated during
remote operation, and cannot be turned on again locally until local operation is
resumed.
Remote command:
*IDN Format
Defines the response format to the remote command
*IDN?
(see
on page 733). This function is intended for re-use of existing control programs
together with the R&S
FSV/A.
"Leg"
Legacy format, as in the R&S FSP/FSU/FSQ family.
"New"
R&S
FSV/A format.
Remote command:
I/O Logging
Activates or deactivates the SCPI error log function. All remote control commands
received by the R&S
FSV/A are recorded in a log file. The files are named according to
the following syntax:
C:\R_S\INSTR\\ScpiLogging\ScpiLog.<no.>
where <no.> is a sequential number
A new log file is started each time logging was stopped and is restarted.
Logging the commands may be extremely useful for debug purposes, e.g. in order to
find misspelled keywords in control programs.
Remote command:
Display Remote Errors
Activates and deactivates the display of errors that occur during remote operation of
the R&S
FSV/A. If activated, the R&S
FSV/A displays a message box at the bottom of
the screen that contains the type of error and the command that caused the error.
The error message remains in place when you switch to "Local" mode. To close the
message box, select the "Close" icon.
Only the most recent error is displayed in remote mode. However, in local mode, all
errors that occurred during remote operation are listed in a separate tab of the "Net-
work + Remote" dialog box (see
Chapter 13.5.5, "Remote Errors"
Network and Remote Control Settings