Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
745
User Manual 1178.8520.02 ─ 01
.............................................................................826
Analyzing Statistics (APD, CCDF)
........................................................................ 838
Measuring the Time Domain Power
......................................................................848
Measuring the Harmonic Distortion
.......................................................................858
Measuring the Third Order Intercept Point
............................................................861
Measuring the AM Modulation Depth
....................................................................864
.................................................................................. 870
14.5.1
Performing Measurements
Useful commands for performing measurements described elsewhere
●
●
Remote commands exclusive for performing measurements:
................................................................................................746
................................................................................ 748
ABORt
This command aborts the measurement in the current channel and resets the trigger
system.
To prevent overlapping execution of the subsequent command before the measure-
ment has been aborted successfully, use the
*OPC?
or
*WAI
command after
ABOR
and
before the next command.
For details see
Chapter 13.1.6.1, "Preventing Overlapping Execution"
Note on blocked remote control programs
:
If a sequential command cannot be completed, for example because a triggered sweep
never receives a trigger, the remote control program will never finish and the remote
channel to the R&S
FSV/A is blocked for further commands. In this case, you must
interrupt processing on the remote channel first in order to abort the measurement.
To do so, send a "Device Clear" command from the control instrument to the
R&S
FSV/A on a parallel channel to clear all currently active remote channels.
Depending on the used interface and protocol, send the following commands:
●
Visa
:
viClear()
Now you can send the
ABORt
command on the remote channel performing the mea-
surement.
Configuring and Performing Measurements