CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult?
<Measurement>
This command queries the results of power measurements.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
irrelevant
<m>
irrelevant
<sb>
Multi-SEM: 1 to 3
for all other measurements: irrelevant
Parameters:
<Measurement>
ACPower | MCACpower
ACLR measurements (also known as adjacent channel power or
multicarrier adjacent channel measurements).
Returns the power for every active transmission and adjacent
channel. The order is:
•
power of the transmission channels
•
power of adjacent channel (lower,upper)
•
power of alternate channels (lower,upper)
The unit of the return values depends on the scaling of the y-
axis:
•
logarithmic scaling returns the power in the current unit
•
linear scaling returns the power in W
CPOWer
Channel power measurements.
Returns the channel power. The unit of the return values
depends on the scaling of the y-axis:
•
logarithmic scaling returns the power in the current unit
•
linear scaling returns the power in W
For SEM measurements, the return value is the channel power
of the reference range (in the specified sub block).
PPOWer
Peak power measurements.
Returns the peak power. The unit of the return values depends
on the scaling of the y-axis:
•
logarithmic scaling returns the power in the current unit
•
linear scaling returns the power in W
For SEM measurements, the return value is the peak power of
the reference range (in the specified sub block).
Note that this result is only available if the power reference type
is set to peak power (see
Configuring and Performing Measurements