Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
761
User Manual 1178.8520.02 ─ 01
CALCulate<n>:LIMit<li>:ACPower:ACHannel:ABSolute
<LowerLimit>[,
<UpperLimit>]
This command defines the absolute limit of the adjacent channels.
If you have defined an absolute limit as well as a relative limit, the R&S
FSV/A uses the
lower value for the limit check.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
Parameters:
<LowerLimit>
The limit of the lower adjacent channel.
Range:
-200 dBm to 200 dBm
*RST:
-200 dBm
Default unit: dBm
<UpperLimit>
The limit of the upper adjacent channel.
Range:
-200 dBm to 200 dBm
*RST:
-200 dBm
Default unit: dBm
Manual operation:
See
See
CALCulate<n>:LIMit<li>:ACPower:ACHannel:ABSolute:STATe
This command turns the absolute limit check for the adjacent channels on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe]
.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
Manual operation:
See
See
CALCulate<n>:LIMit<li>:ACPower:ACHannel[:RELative]
<LowerLimit>[,
<UpperLimit>]
This command defines the relative limit of the adjacent channels. The reference value
for the relative limit is the measured channel power.
If you have defined an absolute limit as well as a relative limit, the R&S
FSV/A uses the
lower value for the limit check.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
Configuring and Performing Measurements