Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
797
User Manual 1178.8520.02 ─ 01
[SENSe:]ESPectrum<sb>:RANGe<ri>:COUNt?
This command queries the number of ranges in the sweep list.
Suffix:
<sb>
.
<ri>
irrelevant
Return values:
<Ranges>
Number of ranges in the sweep list.
Usage:
Query only
[SENSe:]ESPectrum<sb>:RANGe<ri>:DELete
This command removes a range from the sweep list.
Note that
●
you cannot delete the reference range
●
a minimum of three ranges is mandatory.
Suffix:
<sb>
.
<ri>
Selects the measurement range.
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:FILTer:TYPE
<FilterType>
This command selects the filter type for an SEM range.
In case of high speed measurements, the filter has to be identical for all ranges.
Suffix:
<sb>
.
1 | 2 | 3
<ri>
1...30
Selects the measurement range.
Manual operation:
See
[SENSe:]LIST:RANGe<ri>[:FREQuency]:STARt
<Frequency>
This command defines the start frequency of a spurious emission measurement range.
Make sure to set an appropriate span. If you set a span that is
●
smaller than the span the sweep list covers, the R&S
FSV/A will not measure the
ranges that are outside the span - results may be invalid.
●
greater than the span the sweep list covers, the R&S
FSV/A will adjust the start fre-
quency of the first range and the stop frequency of the last range to the span