Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
799
User Manual 1178.8520.02 ─ 01
Parameters:
<Attenuation>
Numeric value.
Refer to the data sheet for the attenuation range.
*RST:
10 dB
Default unit: dB
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:INPut:ATTenuation:AUTO
<State>
This command turns automatic selection of the input attenuation for a SEM range on
and off.
In case of high speed measurements, the input attenuation has to be identical for all
ranges.
Suffix:
<sb>
.
1..n
<ri>
1..n
Selects the measurement range.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
ESP:RANG2:INP:ATT:AUTO OFF
Deactivates the RF attenuation auto mode for range 2.
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:INPut:GAIN[:VALue]
<Gain>
This command selects the gain for a SEM range.
In case of high speed measurements, the level of the preamplifier has to be identical
for all ranges.
Suffix:
<sb>
.
1 | 2 | 3
<ri>
1..30
Selects the measurement range.
[SENSe:]ESPectrum<sb>:RANGe<ri>:INPut:GAIN:STATe
<State>
This command turns the preamplifier for a SEM range on and off.
In case of high speed measurements, the state of the preamplifier has to be identical
for all ranges.
Suffix:
<sb>
.
1..n
Configuring and Performing Measurements