Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
804
User Manual 1178.8520.02 ─ 01
<li>
1..n
Power class for which the limit is defined.
Parameters:
<Level>
Absolute limit at the stop frequency of a SEM range to be used
in addition to the relative limit if the MAX function is enabled (see
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:
RELative:STOP:FUNCtion
Range:
-400 to 400
*RST:
-13
Default unit: dBm
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STOP:
ABSolute -15
Chapter 14.5.6.11, "Example: SEM
Manual operation:
See
" Rel Limit Start / Stop <n> "
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STOP:FUNCtion
<Function>
This command enables the use of a function when defining the relative limit for a SEM
range.
Suffix:
<sb>
.
1..n
<ri>
1..n
Selects the SEM range.
<li>
1..n
Power class for which the limit is defined.
Parameters:
<Function>
OFF | MAX
Defines the function to be used to determine the relative limit
line stop value
MAX
The maximum of the relative
and
the absolute level is used as
RANGe<ri>:LIMit<li>:RELative:STOP
]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:
STOP:ABS
commands to define these values.
OFF
No function is used, the relative limit line is defined by a fixed
relative stop value. Use the
RANGe<ri>:LIMit<li>:RELative:STOP
this value.
*RST:
OFF
Configuring and Performing Measurements