Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
805
User Manual 1178.8520.02 ─ 01
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STOP:
FUNCtion MAX
Chapter 14.5.6.11, "Example: SEM
Manual operation:
See
" Rel Limit Start / Stop <n> "
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:STATe
<State>
This command selects the limit check mode for
all
SEM ranges (<range> is irrelevant).
Suffix:
<sb>
.
1..n
<ri>
1..n
Selects the SEM range.
<li>
1..n
Power class for which the limit is defined.
Parameters:
<State>
ABSolute | RELative | AND | OR
ABSolute
Checks only the absolute limits defined.
RELative
Checks only the relative limits. Relative limits are defined as rel-
ative to the measured power in the reference range.
AND
Combines the absolute and relative limit. The limit check fails
when both limits are violated.
OR
Combines the absolute and relative limit. The limit check fails
when one of the limits is violated.
*RST:
RELative
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:POINts:MINinum[:VALue]
<SweepPoint>
Defines the minimum number of sweep points for the range.
Suffix:
<sb>
.
<ri>
Selects the measurement range.
Parameters:
<SweepPoint>
Minimum number of sweep points per range
Range:
1 to 32001
*RST:
1
Example:
SENSe1:ESPectrum:RANGe3:POINts:MIN:VALue 400
Manual operation:
See
Configuring and Performing Measurements