Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
812
User Manual 1178.8520.02 ─ 01
This command has the same effect as a combination of the
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>[:
commands; however, the power class to be used is not defined directly,
but via the expected power. As opposed to
CALC:LIM:ESP:MODE AUTO
, the power
class is not re-assigned to the input signal power dynamically, but only once when the
command is executed.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<sb>
1..n
Parameters:
<Power>
integer
Range:
-200 to 199
*RST:
0
Example:
CALC:LIM:ESP:VAL 33
Activates manual selection of the limit line and selects the limit
line for P = 33.
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:COUNt
<NoPowerClasses>
This command sets the number of power classes to be defined.
This command must be executed before any new power class values can be defined
using
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:MAXimum
and
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:MINimum
.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<sb>
1..n
<pc>
irrelevant
Parameters:
<NoPowerClasses> 1 to 4
*RST:
1
Example:
CALC:LIM:ESP:PCL:COUN 2
Two power classes can be defined.
Manual operation:
See
" Adding or Removing a Power Class "
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>[:EXCLusive]
<State>
This command selects the power class used by the measurement if
is set to manual.
Note that:
Configuring and Performing Measurements