Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
831
User Manual 1178.8520.02 ─ 01
[SENSe:]LIST:RANGe<ri>:LIMit:STOP
<Level>
This command defines an absolute limit for a spurious emission measurement range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Level>
Absolute limit at the stop frequency of a SEM range.
Range:
-400 to 400
*RST:
13
Default unit: dBm
Manual operation:
See
" Abs Limit Start / Abs Limit Stop "
[SENSe:]LIST:RANGe<ri>:POINts[:VALue]
<Points>
This command defines the number of sweep points in a spurious emission measure-
ment range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Points>
For more information on sweep points see
"How Much Data is Measured: Sweep Points and Sweep
Count"
*RST:
1001
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:RLEVel
<RefLevel>
This command defines the reference level for a spurious emission measurement
range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<RefLevel>
Reference level.
Refer to the data sheet for the reference level range.
*RST:
0 dBm
Default unit: dBm
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:SWEep:TIME
<SweepTime>
This command defines the sweep time for a spurious emission measurement range.
Configuring and Performing Measurements