Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
842
User Manual 1178.8520.02 ─ 01
Suffix:
<t>
.
<gr>
1..n
gate range
Parameters:
<Time>
The value range depends on the gate period you have set for
the selected trace with
•
the stop time may not be higher than the length of the gate
•
the stop time may not be lower than the start time
The reset values depend on the gate range.
•
for gate range 1, the stop time is 1
ms
•
for gate range 3, the stop time is 3
ms
•
for gate range 5, the stop time is 5
ms
Default unit: s
Example:
SWE:EGAT:TRAC1:STOP1 5ms
Sets the stopping point for range 1 on trace 1 at 5 ms.
Manual operation:
See
14.5.8.4
Scaling the Diagram
The following commands set up the diagram for statistical measurements.
CALCulate<n>:STATistics:PRESet
...................................................................................842
CALCulate<n>:STATistics:SCALe:AUTO ONCE
................................................................ 843
CALCulate<n>:STATistics:SCALe:X:RANGe
.....................................................................843
CALCulate<n>:STATistics:SCALe:X:RLEVel
..................................................................... 843
CALCulate<n>:STATistics:SCALe:Y:LOWer
...................................................................... 844
CALCulate<n>:STATistics:SCALe:Y:UNIT
.........................................................................844
CALCulate<n>:STATistics:SCALe:Y:UPPer
.......................................................................844
CALCulate<n>:STATistics:PRESet
This command resets the scale of the diagram (x- and y-axis).
●
Reference level (x-axis)
0.0 dBm
●
Display range (x-axis) for APD measurements
100 dB
●
Display range (x-axis) for CCDF measurements
20 dB
●
Upper limit of the y-axis
1.0
●
Lower limit of the y-axis
1E-6
Suffix:
<n>
.
Configuring and Performing Measurements