CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:AOFF
................................................... 849
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:AVERage
.............................................. 849
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PHOLd
................................................. 850
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary[:STATe]
................................................. 850
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN[:STATe]
.......................................850
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak[:STATe]
...................................... 851
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS[:STATe]
.........................................851
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:SDEViation[:STATe]
................................851
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:AOFF
This command turns all time domain power evaluation modes off.
Suffix:
<n>
.
<m>
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:AVERage
<State>
This command switches on or off averaging for the active power measurement in zero
span in the window specified by the suffix <n>. If activated, a time domain value is cal-
culated from the trace after each sweep; in the end, all values are averaged to calcu-
late the final result.
The number of results required for the calculation of average is defined with
.
Averaging is reset by switching it off and on again.
Synchronization to the end of averaging is only possible in single sweep mode.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:SUMM:AVER ON
Switches on the calculation of average.
AVER:COUN 200
Sets the measurement counter to 200.
INIT;*WAI
Starts a sweep and waits for the end.
Configuring and Performing Measurements