14.5.11.1
Determining the TOI
All remote control commands specific to TOI measurements are described here.
Useful commands for TOI measurements described elsewhere
●
●
CALCulate<n>:DELTamarker<m>:X:RELative?
●
●
●
Remote commands exclusive to TOI measurements
CALCulate<n>:MARKer<m>:FUNCtion:TOI[:STATe]
.......................................................... 862
CALCulate<n>:MARKer<m>:FUNCtion:TOI:SEARchsignal ONCE
...................................... 862
CALCulate<n>:MARKer<m>:FUNCtion:TOI:RESult?
......................................................... 863
CALCulate<n>:MARKer<m>:FUNCtion:TOI[:STATe]
<State>
This command initiates a measurement to determine the third intercept point.
A two-tone signal with equal carrier levels is expected at the RF input of the instru-
ment. Marker 1 and marker 2 (both normal markers) are set to the maximum of the two
signals. Delta marker 3 and delta marker 4 are positioned to the intermodulation prod-
ucts. The delta markers can be modified separately afterwards with the
command.
The third-order intercept is calculated from the level spacing between the normal mark-
ers and the delta markers.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:MARK:FUNC:TOI ON
Switches on the measurement of the third-order intercept.
CALCulate<n>:MARKer<m>:FUNCtion:TOI:SEARchsignal ONCE
This command initiates a search for signals in the current trace to determine the third
intercept point.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Manual operation:
See
Configuring and Performing Measurements