Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
887
User Manual 1178.8520.02 ─ 01
Example:
FREQ:CENT 100 MHz
FREQ:CENT:STEP 10 MHz
FREQ:CENT UP
Sets the center frequency to 110 MHz.
Manual operation:
See
See
[SENSe:]FREQuency:CENTer:STEP
<StepSize>
This command defines the center frequency step size.
You can increase or decrease the center frequency quickly in fixed steps using the
SENS:FREQ UP
AND
SENS:FREQ DOWN
Parameters:
<StepSize>
f
max
is specified in the data sheet.
Range:
1 to fMAX
*RST:
0.1 x span
Default unit: Hz
Example:
//Set the center frequency to 110 MHz.
FREQ:CENT 100 MHz
FREQ:CENT:STEP 10 MHz
FREQ:CENT UP
Manual operation:
See
[SENSe:]FREQuency:CENTer:STEP:AUTO
<State>
This command couples or decouples the center frequency step size to the span.
In time domain (zero span) measurements, the center frequency is coupled to the
RBW.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
FREQ:CENT:STEP:AUTO ON
Activates the coupling of the step size to the span.
[SENSe:]FREQuency:CENTer:STEP:LINK
<CouplingType>
This command couples and decouples the center frequency step size to the span or
the resolution bandwidth.
Parameters:
<CouplingType>
SPAN | RBW | OFF
SPAN
Couples the step size to the span. Available for measurements
in the frequency domain.
Setting Basic Measurement Parameters