Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
897
User Manual 1178.8520.02 ─ 01
[SENSe:]SWEep:TIME
<Time>
This command defines the sweep time. It automatically decouples the time from any
other settings.
Parameters:
<Time>
refer to data sheet
*RST:
depends on current settings (determined automati-
cally)
Default unit: S
Manual operation:
See
See
[SENSe:]SWEep:TIME:AUTO
<State>
This command couples and decouples the sweep time to the span and the resolution
and video bandwidths.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
SWE:TIME:AUTO ON
Activates automatic sweep time.
Manual operation:
See
See
See
[SENSe:]SWEep:TYPE
<Type>
This command selects the sweep type.
Parameters:
<Type>
AUTO
Automatic selection of the sweep type between sweep mode
and FFT.
FFT
FFT mode
*RST:
AUTO
Example:
SWE:TYPE FFT
Manual operation:
See
Setting Basic Measurement Parameters