Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
898
User Manual 1178.8520.02 ─ 01
[SENSe:]SWEep:TYPE:USED
This command queries the sweep type if you have turned on automatic selection of the
sweep type.
Return values:
<Type>
SWE
Normal sweep
FFT
FFT mode
[SENSe:]BWIDth[:RESolution]:FFT
<FilterMode>
[SENSe:]BANDwidth[:RESolution]:FFT
<FilterMode>
Defines the filter mode to be used for FFT filters by defining the partial span size. The
partial span is the span which is covered by one FFT analysis.
This command is only available when using the sweep type "FFT".
Parameters:
<FilterMode>
WIDE | AUTO | NARRow
AUTO
Automatically applies the sweep optimization mode that is best
for the current measurement.
NARRow
Optimizes the sweep mode for a large dynamic range.
WIDE
Optimizes the sweep mode for high performance.
*RST:
AUTO
Example:
BAND:TYPE FFT
Select FFT filter.
Example:
BAND:FFT NARR
Select narrow partial span for FFT filter.
14.7.3
Configuring the Vertical Axis (Amplitude, Scaling)
The following commands are required to configure the amplitude and vertical axis set-
tings in a remote environment.
................................................................................................ 898
...................................................................................900
.....................................................................................903
14.7.3.1
Amplitude Settings
The tasks for manual configuration are described in
Chapter 8.4.2, "Amplitude Set-
Setting Basic Measurement Parameters