Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
919
User Manual 1178.8520.02 ─ 01
Example:
OUTP:TRIG2:PULS:LENG 0.02
Manual operation:
See
14.7.5
Adjusting Settings Automatically
The commands required to adjust settings automatically in a remote environment are
described here.
[SENSe:]ADJust:CONFigure:DURation
............................................................................ 919
[SENSe:]ADJust:CONFigure:DURation:MODE
..................................................................920
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer
.............................................................. 920
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer
............................................................... 920
[SENSe:]ADJust:CONFigure:TRIGger
..............................................................................920
..........................................................................................921
[SENSe:]ADJust:ALL
This command initiates a measurement to determine and set the ideal settings for the
current task automatically (only once for the current measurement).
This includes:
●
Center frequency
●
Reference level
Example:
ADJ:ALL
Manual operation:
See
" Adjusting all Determinable Settings Automatically ( Auto
[SENSe:]ADJust:CONFigure:DURation
<Duration>
In order to determine the ideal reference level, the R&S
FSV/A performs a measure-
ment on the current input data. This command defines the length of the measurement
if
[SENSe:]ADJust:CONFigure:DURation:MODE
MANual
.
Parameters:
<Duration>
Numeric value in seconds
Range:
0.001 to 16000.0
*RST:
0.001
Default unit: s
Example:
ADJ:CONF:DUR:MODE MAN
Selects manual definition of the measurement length.
ADJ:CONF:LEV:DUR 5ms
Length of the measurement is 5 ms.
Manual operation:
See
" Changing the Automatic Measurement Time ( Meastime
Setting Basic Measurement Parameters