Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
935
User Manual 1178.8520.02 ─ 01
SYSTem:COMMunicate:GPIB:RDEVice:GENerator<gen>:ADDRess
<Number>
Changes the IEC/IEEE-bus address of the external generator.
Suffix:
<gen>
.
1..n
Parameters:
<Number>
Range:
0 to 30
*RST:
28
Example:
SYST:COMM:GPIB:RDEV:GEN:ADDR 15
Manual operation:
See
" GPIB Address / TCPIP Address / Computer Name "
SYSTem:COMMunicate:RDEVice:GENerator<gen>:INTerface
<Type>
Defines the interface used for the connection to the external generator.
This command is only available if external generator control is active (see
SOURce<si>:EXTernal<gen>[:STATe]
Suffix:
<gen>
.
Parameters:
<Type>
GPIB | TCPip
*RST:
GPIB
Example:
SYST:COMM:RDEV:GEN:INT TCP
Manual operation:
See
SYSTem:COMMunicate:RDEVice:GENerator<gen>:LINK
<Type>
This command selects the link type of the external generator if the GPIB interface is
used.
The difference between the two GPIB operating modes is the execution speed. While,
during GPIB operation, each frequency to be set is transmitted to the generator sepa-
rately, a whole frequency list can be programmed in one go if the TTL interface is also
used. Frequency switching can then be performed per TTL handshake which results in
considerable speed advantages.
This command is only available if external generator control is active (see
SOURce<si>:EXTernal<gen>[:STATe]
Suffix:
<gen>
.
Parameters:
<Type>
GPIB | TTL
GPIB
GPIB connection without TTL synchronization (for all generators
of other manufacturers and some Rohde & Schwarz devices)