Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
947
User Manual 1178.8520.02 ─ 01
Example:
PMET2:MTIM:AVER ON
Activates manual averaging.
PMET2:MTIM:AVER:COUN 8
Sets the number of readings to 8.
Manual operation:
See
" Average Count ( Number of Readings )"
[SENSe:]PMETer<p>:MTIMe:AVERage[:STATe]
<State>
This command turns averaging for power sensor measurements on and off.
Suffix:
<p>
.
Power sensor index
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
PMET2:MTIM:AVER ON
Activates manual averaging.
Manual operation:
See
[SENSe:]PMETer<p>:ROFFset[:STATe]
<State>
This command includes or excludes the reference level offset of the analyzer for power
sensor measurements.
Suffix:
<p>
.
Power sensor index
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
PMET2:ROFF OFF
Takes no offset into account for the measured power.
Manual operation:
See
[SENSe:]PMETer<p>[:STATe]
<State>
This command turns a power sensor on and off.
Setting Basic Measurement Parameters