Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
950
User Manual 1178.8520.02 ─ 01
Parameters:
<Holdoff>
Time period that has to pass between the trigger event and the
start of the measurement, in case another trigger event occurs.
Range:
0 s to 1 s
Increment: 100 ns
*RST:
0 s
Default unit: S
Example:
PMET2:TRIG:HOLD 0.1
Sets the holdoff time of the trigger to 100 ms
Manual operation:
See
[SENSe:]PMETer<p>:TRIGger:HYSTeresis
<Hysteresis>
This command defines the trigger hysteresis for external power triggers.
The hysteresis in dB is the value the input signal must stay below the IF power trigger
level in order to allow a trigger to start the measurement.
Suffix:
<p>
.
Power sensor index
Parameters:
<Hysteresis>
Range:
3 dB to 50 dB
Increment: 1 dB
*RST:
0 dB
Default unit: DB
Example:
PMET2:TRIG:HYST 10
Sets the hysteresis of the trigger to 10 dB.
Manual operation:
See
[SENSe:]PMETer<p>:TRIGger:LEVel
<Level>
This command defines the trigger level for external power triggers.
Suffix:
<p>
.
Power sensor index
Parameters:
<Level>
-20 to +20 dBm
Range:
-20 dBm to 20 dBm
*RST:
-10 dBm
Default unit: DBM
Example:
PMET2:TRIG:LEV -10 dBm
Sets the level of the trigger
Manual operation:
See
Setting Basic Measurement Parameters