Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
951
User Manual 1178.8520.02 ─ 01
[SENSe:]PMETer<p>:TRIGger:SLOPe
<Edge>
This command selects the trigger condition for external power triggers.
Suffix:
<p>
.
Power sensor index
Parameters:
<Edge>
POSitive
The measurement starts in case the trigger signal shows a posi-
tive edge.
NEGative
The measurement starts in case the trigger signal shows a neg-
ative edge.
*RST:
POSitive
Example:
PMET2:TRIG:SLOP NEG
Manual operation:
See
[SENSe:]PMETer<p>:TRIGger[:STATe]
<State>
This command turns the external power trigger on and off.
Suffix:
<p>
.
Power sensor index
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
PMET2:TRIG ON
Switches the external power trigger on
Manual operation:
See
" Using the power sensor as an external trigger "
14.7.6.5
Configuring the Outputs
The following commands are required to provide output from the R&S
FSV/A.
Configuring trigger input/output is described in
Chapter 14.7.4.3, "Configuring the Trig-
........................................................................................952
......................................................................................... 952
............................................................................................953
Setting Basic Measurement Parameters