DIAGnostic:SERVice:NSOurce
<State>
This command turns the 28 V supply of the BNC connector labeled [noise source
control] on the R&S
FSV/A on and off.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
DIAG:SERV:NSO ON
Manual operation:
See
OUTPut<up>:IF:IFFRequency
<Frequency>
This command defines the frequency for the IF output of the R&S
FSV/A. The IF fre-
quency of the signal is converted accordingly.
This command is available in the time domain and if the IF output is configured for IF.
Suffix:
<up>
.
Parameters:
<Frequency>
*RST:
50.0 MHz
Default unit: HZ
Manual operation:
See
OUTPut<up>:IF:STATe
<State>
Enables or disables output of the measured IF value at the "IF" output connector.
Suffix:
<up>
.
1..n
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0