Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
985
User Manual 1178.8520.02 ─ 01
CALCulate<n>:DELTamarker<m>[:STATe]
<State>
This command turns delta markers on and off.
If necessary, the command activates the delta marker first.
No suffix at DELTamarker turns on delta marker 1.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:DELT2 ON
Turns on delta marker 2.
Manual operation:
See
See
See
CALCulate<n>:DELTamarker<m>:TRACe
<Trace>
This command selects the trace a delta marker is positioned on.
Note that the corresponding trace must have a trace mode other than "Blank".
If necessary, the command activates the marker first.
Suffix:
<n>
.
<m>
Parameters:
<Trace>
Trace number the marker is assigned to.
Example:
CALC:DELT2:TRAC 2
Positions delta marker 2 on trace 2.
CALCulate<n>:DELTamarker<m>:X
<Position>
This command moves a delta marker to a particular coordinate on the x-axis.
If necessary, the command activates the delta marker and positions a reference
marker to the peak power.
Suffix:
<n>
.
Analyzing Measurements (Basics)