Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
991
User Manual 1178.8520.02 ─ 01
ON | 1
Switches the function on
Example:
CALC:MARK:X:SLIM ON
Switches on search limitation.
Manual operation:
See
" Search Limits ( Left / Right )"
See
" Deactivating All Search Limits "
See
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT
<SearchLimit>
This command defines the left limit of the marker search range for
all
markers in
all
windows.
If you perform a measurement in the time domain, this command limits the range of the
trace to be analyzed.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<SearchLimit>
The value range depends on the frequency range or sweep
time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
*RST:
left diagram border
Default unit: HZ
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:LEFT 10MHz
Sets the left limit of the search range to 10 MHz.
Manual operation:
See
" Search Limits ( Left / Right )"
See
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt
<SearchLimit>
This command defines the right limit of the marker search range for
all
markers in
all
windows.
If you perform a measurement in the time domain, this command limits the range of the
trace to be analyzed.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Analyzing Measurements (Basics)