Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
992
User Manual 1178.8520.02 ─ 01
Parameters:
<Limit>
The value range depends on the frequency range or sweep
time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
*RST:
right diagram border
Default unit: HZ
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:RIGH 20MHz
Sets the right limit of the search range to 20 MHz.
Manual operation:
See
" Search Limits ( Left / Right )"
See
CALCulate<n>:MARKer<m>:X:SLIMits:ZOOM[:STATe]
<State>
This command adjusts the marker search range to the zoom area for
all
markers in
all
windows.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:MARK:X:SLIM:ZOOM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:RIGH 20MHz
Sets the right limit of the search range to 20 MHz.
Manual operation:
See
CALCulate<n>:THReshold
<Level>
This command defines a threshold level for the marker peak search (for
all
markers in
all
windows).
Suffix:
<n>
.
irrelevant
Parameters:
<Level>
Numeric value. The value range and unit are variable.
*RST:
-120 dBm
Default unit: DBM