Types of Memory and Information Storage in the R&S
FSV/A
R&S
®
FSVA3000 / R&S
®
FSV3000
7
Instrument Security Procedures 1179.0097.02 ─ 01
4.2
Non-Volatile Memory
The R&S
FSV/A contains no user-accessible non-volatile memory, except for the inter-
nal SSD memory module and the removable CFAST memory card (option R&S
FSV3-
B20, removable hard drive).
All other non-volatile memories of the R&S
FSV/A are not a security concern.
SSD Solid-State Drive (CPU board)
(excludes CFAST memory card R&S
FSV3-B20, removable hard drive)
The internal SSD memory module is located on the CPU board.
Basically, the SSD holds user data and is non-volatile. Hence, user data is not erased
when power is removed from the instrument.
Note:
As you can equip the R&S
FSV/A with the option R&S
FSV3-K33 (security write
protection of solid state drive), the R&S
FSV/A allows you to enable the
. The sanitization procedure depends on the availability of this option:
●
If R&S
FSV3-K33 is
not available
disabled
(default setting
on the instrument):
The R&S
FSV/A saves user data and instrument setups permanently on the SSD.
All data, including user data leaves the secure area with the instrument.
Sanitization procedure:
Sanitizing is not possible without physically removing the
SSD.
●
If R&S
FSV3-K33 is
available
with
enabled
:
The R&S
FSV/A redirects user data and instrument setups to the volatile memory
(SDRAM).
Sanitization procedure:
Turn of instrument power.
See also
Chapter 5, "Instrument Declassification"
Removable CFAST Memory Card (R&S
FSV3-B20, removable hard drive)
(excludes SSD)
Do not remove the CFAST memory card during operation as data can get lost.
●
If the CFAST memory card is
inserted
:
Note:
As you can equip the R&S
FSV/A with the option R&S
FSV3-K33 (security
write protection of solid state drive), the R&S
FSV/A allows you to enable the
. The sanitization procedure depends on the availability of this
option:
–
If R&S
FSV3-K33 is
not available
or
disabled
:
The CFAST memory card holds user data and is non-volatile. User data is not
erased when power is removed from the instrument.
Sanitization procedure:
Remove the CFAST memory card from the instru-
ment.
Non-Volatile Memory