Instrument Declassification
R&S
®
FSVA3000 / R&S
®
FSV3000
8
Instrument Security Procedures 1179.0097.02 ─ 01
–
If R&S
FSV3-K33 is
available
enabled
:
The R&S
FSV/A redirects user data and instrument setups to the volatile mem-
ory (SDRAM).
Sanitization procedure:
Turn of instrument power.
●
If the CFAST memory card is
not inserted
:
Without the CFAST memory card, holding the operating system etc., you cannot
work with the R&S
FSV/A.
Sanitization procedure:
None required.
See also
Chapter 5, "Instrument Declassification"
5
Instrument Declassification
Before you can remove the R&S
FSV/A from a secured area (for example to perform
service or calibration), all classified user data needs to be removed.
Overview for declassification of the R&S
FSV/A depending on the configuration
The options R&S
FSV3-B20 (removable hard drive) and R&S
FSV3-K33 (security write
protection of solid state drive) determine the sanitization procedures for the
R&S
.
Types of Memory and Information Storage in the R&S
, and chapter
Protecting Data Using the Secure User Mode
in the user man-
ual of the R&S
www.rohde-schwarz.com/manual/fsv3000/
illustrates the different ways of data storage and sanitization procedures,
depending on the options the instrument is equipped with.