Trying Out the Instrument
R&S
®
FSVA3000/R&S
®
FSV3000
54
Getting Started 1330.8073.02 ─ 04
To optimize the display
To optimize the display for the calibration signal, we will adjust the main measure-
ment settings.
1. Set the center frequency to the calibration frequency:
a) Tap the "Overview" softkey to display the configuration "Overview".
b) Tap the "Frequency" button.
c) In the "Center" field, enter
64
on the number pad on the front panel.
d) Press the "MHz" key next to the number pad.
2. Reduce the span to 20
MHz:
a) In the "Span" field of the "Frequency" dialog box, enter
20
MHz
.
b) Close the "Frequency" dialog box.
The display of the calibration signal is now improved. The maximum at the
center frequency (=calibration frequency) of 64
MHz becomes visible.
Figure 6-2: Calibration signal with optimized display settings
Measuring a Basic Signal