Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1016
User Manual 1178.8520.02 ─ 01
CALCulate<n>:MARKer<m>:FUNCtion:NOISe:RESult?
This command queries the result of the noise measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Return values:
<NoiseLevel>
Current noise level. The unit is the one currently active.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK2 ON
Switches on marker 2.
CALC:MARK2:FUNC:NOIS ON
Switches on noise measurement for marker 2.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK2:FUNC:NOIS:RES?
Outputs the noise result of marker 2.
Usage:
Query only
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:NOISe[:STATe]
<State>
This command turns the noise measurement at the marker position on and off.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:MARK:FUNC:NOIS ON
Switches on the noise measurement.
Manual operation:
See
See
" Switching All Noise Measurement Off "
Analyzing Measurements (Basics)