Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1042
User Manual 1178.8520.02 ─ 01
<dl>
1 to 4
frequency line
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:FLIN2:STAT ON
Turns frequency line 2 on.
CALCulate<n>:TLINe<dl>
<Time>
This command defines the position of a time line.
Suffix:
<n>
.
<dl>
1 to 4
time line
Parameters:
<Time>
Note that you can not set a time line to a position that is higher
than the current sweep time.
Range:
0 s to 1600 s
*RST:
(STATe to OFF)
Default unit: S
Example:
CALC:TLIN 10ms
Sets the first time line to 10 ms.
Manual operation:
See
CALCulate<n>:TLINe<dl>:STATe
<State>
This command turns a time line on and off
Suffix:
<n>
.
<dl>
1 to 4
time line
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:TLIN:STAT ON
Turns the first time line on.
Analyzing Measurements (Basics)