Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1073
User Manual 1178.8520.02 ─ 01
Secure User Mode
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details, see
Chapter 5.1.7, "Protecting Data Using the Secure User Mode"
Suffix:
<n>
.
irrelevant
Example:
MMEM:STOR:STAT 1,'Save'
Saves the current instrument settings in the file
Save.dfl
.
MMEM:STOR:STAT:NEXT
Saves the current instrument settings in the file
Save_001.dfl
MMEM:STOR:STAT:NEXT
Saves the current instrument settings in the file
Save_002.dfl
Manual operation:
See
MMEMory:STORe<n>:TYPE
<Mode>
This command defines whether the data from the entire instrument or only from the
current channel is stored with the subsequent
MMEM:STOR
... command.
Suffix:
<n>
.
irrelevant
Parameters:
<Mode>
INSTrument | CHANnel
INSTrument
Stores data from the entire instrument.
CHANnel
Stores data from an individual channel.
*RST:
INST
Example:
INST:SEL 'SPECTRUM2'
Selects channel
'SPECTRUM2'
.
MMEM:STOR:TYPE CHAN
Specifies that channel data is to be stored.
SYSTem:PRESet
This command presets the R&S
.
Example:
SYST:PRES
Usage:
Event
Managing Settings and Results