Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1094
User Manual 1178.8520.02 ─ 01
CALibration[:ALL]
<UseData>
This command initiates a calibration (self-alignment) routine and queries if calibration
was successful.
During the acquisition of correction data the instrument does not accept any remote
control commands.
In order to recognize when the acquisition of correction data is completed, the MAV bit
in the status byte can be used. If the associated bit is set in the Service Request
Enable (SRE) register, the instrument generates a service request after the acquisition
of correction data has been completed.
Return values:
<CalState>
ON | OFF | 0 | 1
OFF | 0
Calibration was successful.
ON | 1
Calibration was not successful.
Example:
*CLS
Resets the status management.
*SRE 16
Enables MAV bit in the Service Request Enable register.
*CAL?
Starts the correction data recording, and then a service request
is generated.
Manual operation:
See
CALibration:RESult?
This command returns the results collected during calibration.
Return values:
<CalibrationData>
String containing the calibration data.
Example:
CAL:RES?
would return, e.g.
Total Calibration Status:
PASSED, Date (dd/mm/yyyy): 12/07/2004,
Time: 16:24:54,Runtime: 00.06
Usage:
Query only
Manual operation:
See
DIAGnostic:SERVice:INPut:MC:CFRequency
<Frequency>
This command defines the frequency of the calibration signal for R&S
FSV/A models
43 MHz and higher.
This command only takes effect if a microwave calibration signal is selected for input
(
DIAGnostic:SERVice:INPut[:SELect]
Configuring the R&S
FSV/A