Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1139
User Manual 1178.8520.02 ─ 01
INSTrument:COUPle:CENTer
<State>
This command turns synchronization of the frequency between measurement channels
on and off.
Parameters:
<State>
ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST:
ALL
Example:
INST:COUP:CENT ALL
Synchronizes the center frequency.
INSTrument:COUPle:DEMod
<State>
This command turns synchronization of the audio demodulator configuration between
measurement channels on and off.
Parameters:
<State>
ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST:
NONE
Example:
INST:COUP:DEM ALL
Synchronizes the audio demodulator configuration.
INSTrument:COUPle:GAIN
<State>
This command turns synchronization of the preamplifier configuration between mea-
surement channels on and off.
Parameters:
<State>
ALL | NONE
ALL | 1
Turns on synchronization.
NONE | 0
Turns off synchronization.
*RST:
NONE
Example:
INST:COUP:GAIN ALL
Synchronizes the preamplifier configuration.
Configuring the R&S
FSV/A