Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1144
User Manual 1178.8520.02 ─ 01
*RST:
1
Example:
INST:COUP:USER3 'Spectrum1','All
Windows','Attenuation','Spectrum 2','All
Windows','Attenuation',BID,ON
Synchronizes the attenuation between the channels named
'Spectrum1' and 'Spectrum 2' in both directions and turns on the
coupling.
Manual operation:
See
INSTrument:COUPle:USER<uc>:CHANnel:LIST?
This command queries the names of the measurement channels that can be synchron-
ized.
Suffix:
<uc>
.
Index of a user-defined parameter coupling. Which indexes are
available is defined by
Return values:
<SynchronizableChannel>
Comma-separated list of strings
All channels that can be synchronized.
Example:
INST:COUP:USER:CHAN:LIST?
Result:
'SPEC1','AD1','All Spectrum','All
Channels','All Analog Demod'
Usage:
Query only
Manual operation:
See
INSTrument:COUPle:USER<uc>:ELEMent:LIST?
[<ChannelName>, <Parameter>]
This command queries parameters that can be synchronized.
Suffix:
<uc>
.
irrelevant
Query parameters:
<ChannelName>
Optional SCPI parameter.
String containing the name of a measurement channel.
<Parameter>
Optional SCPI parameter.
String containing the name of a parameter that you can syn-
chronize.
Return values:
<SynchronizableParameter>
Comma-separated list of parameters.
No parameters provided
Parameters that can be synchronized for all channels
Configuring the R&S
FSV/A