Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1145
User Manual 1178.8520.02 ─ 01
Channel name provided
Parameters that can be synchronized for the selected channel
Parameter and channel name provided
Parameters that can be synchronized with the specified parame-
ter for the selected channel.
Example:
INST:COUP:USER:ELEM:LIST?
Result: all parameters that can be coupled:
'AC DC Coupling', 'Attenuation', 'Center
Frequency', 'Display Lines', 'Frequency Marker
1',...
Example:
INST:COUP:USER:ELEM:LIST? 'Spectrum'
Result: all parameters that can be coupled in the 'Spectrum'
channel:
'AC DC Coupling','Attenuation','Center
Frequency','Display Lines',...
Example:
INST:COUP:USER:ELEM:LIST? 'Spectrum',
'Attenuation'
Result: all parameters that can be coupled to 'Attenuation' in the
'Spectrum' channel:
'Attenuation'
(Attenuation is the only parameter that can be coupled to attenu-
ation.)
Usage:
Query only
Manual operation:
See
"Coupling Element 1 / Coupling Element 2"
INSTrument:COUPle:USER<uc>:INFO
This command queries additional information about the specified user-defined parame-
ter coupling.
Suffix:
<uc>
.
Index of a user-defined parameter coupling. Which indexes are
available is defined by
Return values:
<Information>
String containing the message as displayed in the coupling man-
ager.
If the coupling message contains no message, an empty string
is returned.
Example:
INST:COUP:USER2:INFO?
Queries possible information about the user coupling with index
2.
Result:
'Only one limit line allowed'
Manual operation:
See
Configuring the R&S
FSV/A